Sensor/Transmitter, Toxic Gas & Oxygen

ENMET Corporation

15


Phosgene (3E sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

ammonia

50

5, drops to 0 in short time

abundant change in humidity

yes

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

hydrogen sulfide

1*

sulfur dioxide

2

hydrochloric acid

5*

carbon monoxide

300

carbon dioxide

5000

nitrogen

100%

chlorine

1

hydrocarobons

% range

hydrogen

1000

*With onboard filter; continuous high level exposure may reduce the filter efficiency

Silane (3E 50 sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

chlorine

0,5

-4

hydrogen cyanide

10

1

ammonia

100

<1

diborane

100

0.4

arsine

1

0.7

germane

1

1

disilane

5

yes

phosphine

300ppb

0.2

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

hydrogen sulfide

1*

sulfur dioxide

2

hydrochloric acid

5

carbon monoxide

300

carbon dioxide

5000

nitrogen

100%

chlorine

1,5

hydrocarobons

% range

hydrogen

1000

hydrogen fluoride

3.8

hydrogen selenide

50ppb

*With onboard filter; continuous high level exposure may reduce the filter efficiency
‡Short term gas exposure (min.)

Notes: