Sensor/Transmitter, Toxic Gas & Oxygen

ENMET Corporation

10


Hydrogen Cyanide(2E 30 sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

hydrogen sulfide

15

40

chlorine

5

-1*

nitrogen dioxide

15

-40

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

hydrogen chloride

10*

methane

2000

carbon monoxide

300

carbon dioxide

5,000

gasoline

300

sulfur dioxide

50*

nitric oxide

100

chlorinated HC's

<200

†Long term exposure may destroy sensor
*with onboard filter; continuous high level exposure may reduce the filter efficiency