Chapter three – Heath Consultants Detecto-Pak 4 (DP) User Manual

Page 17

Advertising
background image

100786-0 Rev. D

05/09

10

CHAPTER THREE

III.

CALIBRATION MOTHER TANKS/CALIBRATION

CYLINDERS:

This Chapter discusses, in detail, the proper steps to take in transferring

calibration gas to a calibration cylinder.

Recommendations for ordering calibration gas mother tanks, along with

proper storage, are also discussed.

A. CALIBRATION GAS MIXTURES

The calibration gas is comprised of a mixture of 100 ppm + 10% methane,

balance zero air. The calibration gas mixture is drawn into the instrument

at the inlet of the filter housing to ensure proper sensitivity level (See

Chapter Six for a detailed explanation of this method.)

B. TRANSFILLING CALIBRATION CYLINDERS

Heath assumes no liability in the refilling of calibration cylinders. If you

wish to refill your own calibration cylinders, caution must be exercised

during the refilling process. It is also recommended that all safety precau-

tions be used in the handling of high-pressure compressed, flammable

gases during the refilling process.

To refill the calibration cylinder, connect the calibration transfiller to

a calibration mother tank, then attach a non contaminated calibration

cylinder to the Transfiller. See Figure 3 this Section and **NOTE**,

Section II, C of Chapter Two.