Select the 10 ppm range and fine zero the instrument using the

zero controls on the top panel. Press and release a zero control to

make adjustments. Wait at least one second between zero control

presses.

**NOTE**

While in operation, it is normal for the zero to drift with ambient

temperatures and small changes in the methane background concentration.

The fine zero controls also support a “fast” mode equal to 10 single presses.

Fast mode is enabled by holding down either zero control for longer then

one second. Every second thereafter will produce the equivalent to 10 single

presses followed by a beep. Release the zero control to exit the fast mode.

The Detecto-Pak 4 should now be zeroed to within 1 ppm.

D. SHUTDOWN PROCEDURES

The following procedure must be followed in order to maintain the

Detecto-Pak 4 in good working order. This procedure should be followed

whenever the Detecto-Pak 4 remains inactive for two or more days. The

battery charging procedure is all that will change.

**NOTE**

The Detecto-Pak 4 should only be shut down and the battery changed or

charged in a nonflammable (clean air) atmosphere. Instrument should be

in 10K range.

1. Disconnect and remove the fuel cylinder from the instrument. Install

the protective cap on the fuel cylinder valve and store the cylinder in

its proper compartment in the carrying case.

2. Wait for the fuel to purge from the system tubing. The continuous

“flame out” alarm will sound when the fuel is purged and the flame

goes out.

3. Press the Power function switch to turn the Detecto-Pak 4’s electronics

off.

4. Disconnect the probe and store the instrument and probe in their

proper compartments in the carrying case, or connect the instrument to

its automatic battery charger. (See Chapter Four for battery charging

procedures.)