Heath Consultants Detecto-Pak 4 (DP) User Manual

Page 33

Advertising
background image

CALIBRATION

100786-0 Rev. D

05/09

26

IMPORTANT

Special procedures must be followed when using the Detecto-Pak 4 at high

altitudes (over 3,000 feet above sea level). A slightly richer fuel mixture

(42% hydrogen/58% nitrogen) and increased fuel and sample flow rate

(50-55 cc/m and 850 cc/m, respectively) are required. The instrument must

be calibrated at elevation and re-calibrated if the elevation changes by

1000 feet or more. DO NOT use these settings below 3000 feet as the flame

ionization detector (FID) will burn too hot and possibly cause operational

and/or physical problems. If an improper (hydrogen higher than 42%) fuel

mix is used, the possibility of damage to the detector cell could be realized

as excessive heat is generated. Please contact servicing Heath Office for

proper ordering procedures.

9.

The meter needle should register 100 (or same as calibration gas

used) on the 100 ppm range. If the needle is not on 100, you will

need to reset the instrument with the calibration (RP6) port.

10. If an adjustment is necessary to bring the meter needle to 100

(or same as calibration gas used) on the 100 ppm meter scale, an

adjusting control is inside the bottom rear of the instrument as

indicated in Figures 7 & 8. Access the CALIBRATION adjustment

through the LEFT-HAND ACCESS PORT on the rear of the

instrument located directly below the fuel cylinder holder.

11. Remove the plug from the port and use a micro-screwdriver to

adjust the CALIBRATION level using RP6 as required.

12. Repeat the above steps to verify the ZERO and CALIBRATION

levels.

13. Disconnect the calibration hose fitting from the sample inlet quick

disconnect on the filter housing and close the calibration cylinder

valve. Calibration is now complete.

Advertising