15. SIGNAL ALARM: Audible alarm used to indicate the various types

of alarm conditions which may exist. A continuous alarm is used

for flame out, low sample flow, or in the rare event of detector cell

over-heating. A pulsating alarm is used for indicating the signal level

exceeds the alarm set point.

16. Please use Heath supplied charger only to charge DP4 through this

charger jack.

**NOTE**

The instrument will verify all control switch operations by making a short

beep. This is normal and does not indicate an alarm. Alarms typically last

several seconds or as long as the condition lasts.

B. OPERATING PROCEDURES

The Detecto-Pak 4 is easy to operate. It is important that the following

procedure be carefully read and followed in order to obtain the maximum

benefits of its many features.

**NOTE**

The Detecto-Pak 4 should only be turned on and zero adjusted in a

nonflammable (clean air) atmosphere.

1. Carefully unpack all items from the system carrying case.

2. Always start with a fully charged battery. (See Chapter Four, Section

IV, A-D)

3. Always start with enough fuel in the fuel cylinder to last the expected

operating time. 1,750 psig in the two (2) fuel cylinders will ensure

at least eight hours of operation. Do not allow pressure in the fuel

cylinder to drop below 50 psig.

**NOTE**

Fuel cylinders with negative pressure (vacuum) are supplied on the initial

shipment. The operator must fill cylinders to operating fuel level before

start up of the Detecto-Pak 4.

**NOTE**

Apply a very small amount of supplied “O” ring lubricant to small “O” ring

on the stem of the cylinder valve at least once a week.