B. BATTERY CHARGER

It is very important to use ONLY the approved charger supplied with

the Detecto-Pak 4 to recharge the battery pack. Substitute chargers may

damage the battery pack. After the charger has been plugged into the

Detecto-Pak 4 for 10 hours, the charger will automatically reduce itself

to the float mode. The float mode will maintain a charge in the battery as

long as it is connected to the charger. We recommend the Detecto-Pak

4 be connected to the charger whenever it is not it use. This will ensure

the battery is charged and the instrument is ready for use.

C. CHARGING PROCEDURE

WARNING

THE BATTERY MUST BE CHANGED OR CHARGED IN AN AREA

KNOWN TO BE NONFLAMMABLE TO REDUCE THE RISK OF

IGNITING A FLAMMABLE ATMOSPHERE.

The charging procedure is to set the Detecto-Pak 4 instrument Power

Switch to “Off.” Plug the battery charger into a 120 VAC receptacle,

then plug the battery charger output plug into the charger jack on the

Detecto-Pak 4. If connected in this correct sequence, the charger is on

and in the high charge mode.

Charging time is normally a 10 to 12 hour period if the instrument has

been operated for 8 to 10 hours. Although shorter charging times are not

recommended, if it becomes necessary use a 1.5 to 1 ratio to time in use.

Example: If you use the instrument for 4 hours, charge the battery for 6

plus hours to completely recharge it.

D.

GENERAL INFORMATION - NICKEL METAL HYDRIDE

BATTERY

1. We recommend that a new replacement Nickel Metal Hydride battery

by cycled 2 times before regular use. First, fully charge the battery,

then discharge the battery by operating the instrument for 8 to 10

hours or preferably until it ceases functioning. Then fully recharge

for 10 plus hours. This is not necessary for new instruments shipped

from the factory and put into use within 2 weeks because the cycling

was accomplished in the factory.