VI. CALIBRATION:

The Heath Detecto-Pak 4 is factory calibrated. It is recommended that the

Detecto-Pak 4 be calibrated once a week. However, it is recognized that

the Detecto-Pak 4 will be used for a wide range of applications and for

varying lengths of time. Such variables may cause the operator to develop

his or her own calibration schedule. The operator may see a need for a

more or less frequent calibration schedule.

The Detecto-Pak 4 should also be calibrated after any repair or replacement

of any parts and at any time the operator suspects the instrument’s

calibration has changed.

In addition to calibrating the instrument on a regular basis, the instrument

should be run through a diagnostic check before each use. The check

should include an observation of the external sampling system. Also, a

daily operational check, with a known sample of gas is recommended to

insure the instrument is operating.

A. CALIBRATION GAS

The type of calibration gas and the concentration level of the

calibration gas will be dictated by the particular detection needs

requirements.

The most reliable detection results will be obtained when the Detecto-

Pak 4 is calibrated with a calibration gas containing the same gas

component anticipated in the test area.

Additionally, the concentration of the calibration gas should represent

the level nearest the critical area of detection required. That is, if

an anticipated detection level of a particular gas is 100 ppm, then

a calibration gas of that species at a 100 ppm + 10% concentration

level would be the best calibration gas to use.

Methane has been accepted in the industry as a general calibration

standard for flame ionization detectors such as the Detecto-Pak 4. It

is recommended the Detecto-Pak 4 be calibrated with 100 ppm of

methane + 10% in air for general detecting applications.

**NOTE**

The calibration of the Detecto-Pak 4 with a particular gas species will

provide reliable ppm readings for a target gas of the same species. Meter

readings observed from the testing of a target gas containing a gas

different from the calibration gas species may not represent the actual ppm

concentration in the sampled target gas.