CALIBRATION

100786-0 Rev. D

05/09

25

7. Attach calibration kit regulator assembly to calibration cylinder;

turn calibration cylinder valve knob counterclockwise (one-half

to one turn). DO NOT FORCE THE KNOB ALL THE WAY

COUTERCLOCKWISE (OPEN)

8. Attach calibration kit outlet connector to the portable unit by

pushing the connector firmly into the sample inlet quick connector

on the filter housing. Within 30 seconds, the meter needle should

start moving upscale. Allow another one-half minute (30 seconds)

before checking the meter.

(Figure 6)