Chapter 2: Prerequisites

Product Service 1-800-522-7658

6

Original Instructions


All Purifier Clean Benches with model numbers ending in -00, -01, -20 & -21 are
designed for operation at 115 volts, 60 Hz, alternating current. Clean Benches
with model numbers ending in -02 through -12 & -22 through -32 are designed for
operation at 230 +/- 20 volts, 50/60 Hz alternating current. A dedicated outlet
with a circuit breaker rated at 20 amps should be located as close as possible to
the right rear side of the bench, at a height even with, or higher than, the bottom
of the bench in its final location. If your electrical outlet is distant from this
location, contact Labconco’s Product Service Department for information on
longer cords.

Do not use any detachable power cord that is not adequately
rated for the unit.

Ne pas utliser un fil électrique amovible qui n’est pas du tension
nominale de l’appareil.


Service Line Requirements


All service lines to the Clean Bench should be quarter inch outside diameter,
metal, and equipped with an easily accessible shut-off valve should disconnection
be required. If the service line pressure exceeds 40 PSI, it must be equipped with
a pressure regulator to reduce the line pressure.

The use of flammable gases or solvents should be avoided in
the clean bench. Open flames in the cabinet will disrupt the
laminar airflow in the bench. If you feel that your procedure
requires the use of an open flame or flammable materials,
contact the appropriate safety official.

The use of air or gases under high pressure should be
considered carefully as they may seriously disrupt the airflow
patterns in the bench.

Model #

Requirements

Nominal Width

3889700, -01, -20, -21

115 VAC, 60 Hz, 7 Amps

3'

3889702 to -12 & -22 to -32

230 VAC, 50/60 Hz, 4 Amps

3'

3888400, -01, -20, -21

115 VAC, 60 Hz, 7 Amps

4'

3888402 to -12 & -22 to -32

230 VAC, 50/60 Hz, 4 Amps

4'

3250000, -01, -20, -21

115 VAC, 60 Hz, 10 Amps

5'

3250002 to -12 & -22 to -32

230 VAC, 50/60 Hz, 5 Amps

5'

3873000, -01, -20, -21

115 VAC, 60 Hz, 10 Amps

6'

3873002 to -12 & -22 to -32

230 VAC, 50/60 Hz, 5 Amps

6'

3818500, -01, -20, -21

115 VAC, 60 Hz, 16 Amps

8'

3818502 to -12 & -22 to -32

230 VAC, 50/60 Hz, 8 Amps

8'

!

Table 2-1

!

!