background image

Chapter 6: Maintaining Your Clean Bench

Product Service 1-800-522-7658

27

Original Instructions


Changing the Fluorescent Lamp:

1. Unplug the Purifier Clean Bench.
2. Remove the prefilters as described earlier in this chapter.
3. Remove the front dress panel by removing the four screws located on the top

front corners of the panel, and then lifting the panel up and forward.

4. Remove the lamp reflector by gently squeezing the reflector to compress it

and then lift the reflector straight up.

5. Remove the lamp by rotating it out of its sockets and pulling it straight up.
6. Make sure that the surface of the new lamp is clean and free of dirt or

fingerprints, by cleaning its surface with alcohol or glass cleaner.

7. Install the new lamp by reversing the above steps.

Changing the UV Lamp (if option is installed):

1. Unplug the Purifier Clean Bench.
2. Access to the UV Lamp is from within the Clean Bench work area, the bulb is

located on the roof of the cabinet in the work area.

3. (On 3 & 4 Foot models) Remove the UV Lamp by rotating it out of its sockets

and pulling it down.

3. (On 6 & 8 Foot models) Remove the UV Lamp by sliding the bulb axially
toward the side of the unit and then pull down.
4. Make sure that the surface of the new UV Lamp is clean and free of dirt or

fingerprints, by cleaning its surface with alcohol or glass cleaner.

5. Install the new UV Lamp by reversing the above steps.

The service operations listed in the rest of this chapter should
only be performed by a qualified certifier.

Les operations de service listées dans le reste de ce chapitre ne
doit être effectuée que par un certificateur qualifié.



Speed Control Adjustment:

Adjusting the speed control will have an effect on the air
velocities and the effectiveness of the clean bench. Only a
qualified certification technician, as part of the recertification
process, should adjust the speed control.

Do NOT contact blower wheel while still in motion.
NE PAS être en contact avec la roué du ventilateur tant qu’il est
en marche.

!

!

!