Refer to the following table if your Purifier Clean Bench fails to operate properly.
If the suggested corrective actions do not solve your problem, contact Labconco
Product Service for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Blower and lights
won’t turn on

Unit not plugged into
outlet

Plug the clean bench into appropriate
electrical service.

Circuit breakers
tripped

Reset circuit breakers.

Lights work but
motor won’t run

Speed control out of
adjustment

Have speed control adjusted by a
qualified certifier.

Defective speed
control

Replace speed control.

Defective motor

Replace motor.

Pressure gauge
reading increases
(more than 0.1 in.)

Blockage of the
prefilter or diffuser

Ensure that both the prefilter and
diffuser are clear.

HEPA filter loading

The gauge reading will increase when
the HEPA filter is plugged.

Contamination of
work in the clean
bench

Improper technique
or procedure for the
clean bench

See “Use of the Clean Bench” section
of this manual.