Chapter 6: Maintaining Your Clean Bench

Product Service 1-800-522-7658

32

Original Instructions

5. A finer adjustment can be made to the shade closing height by backing the

two torx head (T20) screws out one or two turns on each of the UV Shade
support brackets, in order to allow the support brackets to slide up or down
one hole notch on the brackets. The UV Shade Cover does not need to be
removed for this adjustment.

`




Adjusting the UV Shade Sensor Bracket:

Should the need arise to adjust the position of the Sensor Bracket on the UV
Shade handle to ensure proper contact with the UV Lockout Sensor at the far right
end of the Worksurface, perform the following steps as necessary.

1. Lower the UV Shade to the normal closed height, where the UV Shade

Handle is resting at the Stainless Steel Worksurface.

Loosen but do NOT
remove these 2 screws
(and same 2 screws on
opposite bracket) for
fine adjustment

To remove UV Shade
Cover, remove this
screw (and same screw
on opposite end)

Bend bracket away
from UV Shade slightly
to pull tab out of slot

Rotate Tab with pliers one
ratchet click up or down, if
necessary

Crossed slots allow Tab to
be inserted in any of the 90
degree positions for
additional adjustment

Figure 6-5

Figure 6-6