Operating the uv light – Labconco Purifie Horizontal Clean Benches 38185 Series, 8' User Manual

Page 27

Advertising
background image

Chapter 5: Using Your Clean Bench

Product Service 1-800-522-7658

23

Original Instructions

• Avoid the use of an open flame.

• Use proper aseptic technique. – “Sterile Air First”

• Avoid using techniques or procedures that disrupt the airflow patterns of

the clean bench.

Final Purging

• Upon completion of work, the clean bench should be allowed to operate

for 2 to 3 minutes undisturbed, to purge airborne contaminants from the
work area.

Wipe-Down

• Wipe down the interior surfaces of the clean bench with stainless steel

cleaner or 70% ethanol and allow to dry.

Shutdown

• Turn off the fluorescent light and clean bench blower.

Operating the UV Light


To activate the ultraviolet light, the following steps must be taken:

1. Turn the blower switch to the “Off” position, as shown in Figure 5-1.
2. Lower the UV Shade to the worksurface.
3. Select the desired length of time by pressing the Timer Button repeatedly

until the appropriate value is selected. Selecting the Infinity “∞” value
will activate the UV light until you manually deactivate it.

4. Set the Light Switch to the “UV” position, as shown in Figure 5-1.


U.V. Light is restricted by interlock. Blower Switch must be in
off position for U.V. Light operation. Avoid direct exposure of
eyes and skin to ultraviolet light.

La lumière ultrviolette est interdit par verrouillage.
L’interrupteur du ventilateur doit être dans la position d’arrête
pour le fonctionnement de la lampe. Evitez toute exposition
directe aux yeux et à la peau de la lumière ultraviolette.


!

!

Advertising