Chapter 3: Getting Started

Product Service 1-800-522-7658

14

Original Instructions

The clean bench is very top heavy! Use caution when lifting or
moving it.


L’hotte est très lourde du haut! Soulever ou déplacer avec
précaution.


4. Carefully lift the clean bench onto the top rails of the base stand and slide it

into position. The rear of the bench should be flush with the rear of the base
stand and the sides of the clean bench will overhang the side rails of the base
stand approximately 0.25 inches on both sides.

5. Insert the 2.5 inch x 0.25 inch mounting bolts through the hole in the

underside of the base stand and secure it with a flatwasher and lockwasher.

6. Repeat steps 5 and 6 for the remaining bolts.
7. Tighten all four bolts securely.

Initial Certification


Prior to use, all Purifier Clean Benches should be certified by a qualified certifier.
Under normal operating conditions, the Purifier Clean Bench should be recertified
at least annually and when moved or serviced. The certifier should perform the
following tests, as recommended in Institute of Environmental Sciences and
Technology, IEST RP-CC0002.2:

• Airflow Velocity Test

• HEPA Filter Leak Test

• Introduction Leak Test/Backstreaming (when appropriate)

• Airborne Particle Count (when appropriate)

• Lighting Intensity Test (when appropriate)

• Noise Level Test (when appropriate)

• Vibration Test (when appropriate)


In addition, the following tests should also be performed at the user’s discretion:

• Electrical Leakage and Ground Circuit Resistance Test

• Measurement of Line Voltage and Current

• Smoke Test to determine airflow patterns


If you have any questions regarding certification agencies or need assistance in
locating one, contact Labconco’s Product Service Department at 1-800-522-7658
or 816-333-8811.

!

!