Chapter 6: Maintaining Your Clean Bench

Product Service 1-800-522-7658

26

Original Instructions

Service Operations


Resetting the Circuit Breaker:

1. The circuit breaker is located next to the power cord on the upper rear panel of

the bench as shown in Figure 6-1. If the circuit breaker trips, it can be reset
by pressing the white button in.



Changing the Prefilter:

The prefilter should be replaced at least quarterly, or more often as conditions
require. See Appendix A: Clean Bench Components for replacement prefilter
ordering information.

1. Make sure the clean bench is off.
2. Using a Phillips screwdriver, remove the prefilter retainer(s) located on the

front edge of the prefilters, as shown on Figure 6-1.

3. Remove the center prefilter by lifting the front edge straight up and forward.
4. Remove the side prefilters by sliding them towards the center of the bench,

then lifting the front edges straight up and forward.

5. Install new prefilters by reversing the above steps.

Do NOT contact blower wheel while still in motion.

NE PAS être en contact avec la roué du ventilateur tant qu’il est
en marche.

Figure 6-1

Prefilter Retainers
Loosen and remove both screws
to remove each retainer.

Circuit Breaker
(push to reset)

!