Chapter 4: Theory Of Operation And Safety Precautions


HEPA filters are only effective for the
entrapment of particulate matter.
Manipulations which generate gases
or vapors, i.e., toxic chemicals or
radionuclides, should not be
performed in this station.

The PuriCare Vertical Flow Station
should be certified by a qualified
certification technician before its
initial use. The station should be
recertified whenever it is relocated,
serviced or at least annually
thereafter.

Some components of the PuriCare
Vertical Flow Station should only be
serviced by a qualified certification
technician. Some internal components
of the PuriCare may become
contaminated with allergenic
materials during operation of the unit.
Only experienced personnel
competent in decontamination
procedures should decontaminate the
station before servicing contaminated
components. If you have any
questions regarding certification
agencies, or need assistance in locating
one, contact Labconco’s Product
Service Department at 800-522-7658
or 816-333-8811.

Ensure that the unit is connected to
electrical service in accordance with
local and national electrical codes.
Failure to do so may create a fire or
electrical hazard. Do not remove or
service any electrical components
without first disconnecting the
PuriCare from electrical service.

!

!