Service operations – Labconco PuriCare Vertical Flow Stations 3830020 User Manual

Page 31

Advertising
background image

27

Product Service 1-800-522-7658

Chapter 6: Maintaining Your PuriCare

• Using an appropriate optical glass cleaner and

cloth, clean the sash.

• Operate the station blowers, noting the pressure

readings in an operational log.

• Lift the work surface and inspect the prefilters.


Monthly (or more often as required)

• Using a damp cloth, clean the exterior surfaces

of the station, particularly the front, back and
top of the unit, to remove any accumulated dust.

• Disinfect and lift the work surface. Surface

disinfect the lower plenum with a solution of
70% ethanol, or a suitable disinfectant. Check
the prefilters for retained materials.

• Replace the prefilters as needed.

• All weekly activities.


Annually

• Have the station recertified by a qualified

certification technician.

• All monthly activities.


Biannually

• Replace the fluorescent lamp.

Service Operations


Work Surface Removal:

1. Loosen and remove the three thumbscrews located

at the front of the work surface. Only loosen the
three thumbscrews on the back of the work surface.

2. Lift the front edge of the work surface up and pull it

out of the station work area.


Prefilter Removal and Cleaning:

The prefilters may be contaminated
with allergenic material. Take
appropriate precautions when
removing and handling the prefilters
to prevent exposure. The prefilters
are disposable and cannot be cleaned.

!