Refer to the following table if your Vertical Flow
Station fails to operate properly. If the suggested
corrective actions do not solve your problem, contact
Labconco for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Station blower and
lights won’t turn on

Unit not plugged into
outlet

Plug the PuriCare into appropriate
electrical service.

Circuit

breakers

tripped

Reset circuit breakers.

Blower(s) won’t
turn on but lights
work

Blower(s) wiring is
disconnected

Inspect blower(s) wiring.

Blower(s)

motor

is

defective

Replace blower(s) motor.

Station blower turns
on but lights don’t
work

Lamp(s) not installed
correctly

Inspect lamp installation.

Lamp is defective

Replace lamp.

Lamp wiring is
disconnected

Inspect lamp wiring.

Defective

lamp

ballast

Replace lamp ballast.