Refer to the following table if your Vertical Flow
Station fails to operate properly. If the suggested
corrective actions do not solve your problem, contact
Labconco for additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Station blower and
lights won’t turn on
Unit not plugged into
outlet
Plug the PuriCare into appropriate
electrical service.
Circuit
breakers
tripped
Reset circuit breakers.
Blower(s) won’t
turn on but lights
work
Blower(s) wiring is
disconnected
Inspect blower(s) wiring.
Blower(s)
motor
is
defective
Replace blower(s) motor.
Station blower turns
on but lights don’t
work
Lamp(s) not installed
correctly
Inspect lamp installation.
Lamp is defective
Replace lamp.
Lamp wiring is
disconnected
Inspect lamp wiring.
Defective
lamp
ballast
Replace lamp ballast.