Labconco PuriCare Vertical Flow Stations 3830020 User Manual

Page 47

Advertising
background image

43

Product Service 1-800-522-7658

Chapter 8: Troubleshooting


PROBLEM

CAUSE

CORRECTIVE ACTION

Slight increase in
pressure reading

HEPA filter loading

The pressure reading will steadily
increase as the unit is used.

Prefilter loading

Check the prefilters for obstruction(s).
Clean if necessary.

Blockage

or

restriction under the
work surface

Ensure that the plenum beneath the
work surface is unobstructed.

Blockage of the
exhaust outlet

Ensure that the exhaust outlet is not
blocked or restricted.

Contamination of
work in the station

Improper technique
or procedure for the
Vertical Flow Station

See “Use of the Vertical Flow Station”
section in the manual.

Restriction of the
grille – blockage of
the exhaust outlet

Ensure that all return air slots, grilles
and the exhaust outlet are unobstructed.

External factors are
disrupting the
station’s airflow
patterns or acting as a
source of
contamination

See “Installation” section of this
manual.

Station is out of
adjustment/HEPA
filter(s) are defective

Have station recertified.