Chapter 6: Maintaining Your PuriCare
Supply HEPA Filter
Replacement
The HEPA filters should only be
serviced by a qualified certification
technician. Following replacement of a
HEPA filter, the station MUST be
recertified by a qualified certification
technician.
1. Unplug the station.
2. Remove the two upper prefilters as described earlier
in this chapter.
3. Remove the five screws that secure the upper rear
panel as shown in Figure 6-5.
4. Carefully swing the rear panel out.
!
Figure 6-4