Model 38300

Sash Height (inches)

14

Downflow (FPM)

50-55

Approximate Settings for new Filters

Mag Gauge Reading

0.3-0.4 upper / 0.5-0.7 lower

Supply Air Volume Displacement (CFM)

440

Number of Laskin Nozzles Needed

1

Exhaust Air Volume Displacement (CFM)

650-670

Number of Laskin Nozzles Needed

2

Supply HEPA Filter Dims. (in.)

48x18x3

Exhaust HEPA Filter Dims (in.)

48x18x6

Fluorescent Light(s)

1 each F32T8/TL741


*For further details on the secondary inflow method,
consult the inflow velocity calculation – Secondary
Method section of Chapter 6: Maintaining Your
PuriCare
.

Use the downflow velocity test grid as described in this
manual when checking the downflow air velocity.