Inflow volume calculation - secondary method – Labconco PuriCare Vertical Flow Stations 3830020 User Manual

Page 35

Advertising
background image

31

Product Service 1-800-522-7658

Chapter 6: Maintaining Your PuriCare


5. Place the pattern with the hole over the other

prefilter hole. Seal the pattern to the prefilter shelf
if necessary.

6. Place the DIM upside down over the hole in the

second pattern. Seal the DIM to the pattern with
duct tape if necessary.

7. Start the Station and let it run for at least two

minutes. Turn on the DIM and take at least 10
readings of the inflow volume. Average the
readings.

8. The average inflow volume, corrected for local

conditions, should be 650-670 CFM.



Inflow Volume Calculation -
Secondary Method


1. Turn the station off and raise the sashes.
2. Remove the work surface and the lower prefilters as

described earlier.

3. Using rigid cardboard or fiberboard, cut a 20 inch

square.

4. Use the square to cover the prefilter hole directly

above the blower.

5. At the other prefilter hole, mark a test grid at 4 ½

inch intervals around the perimeter of the hole,
establishing a 16 square grid in the hold.

Figure 6-3