38

Product Service 1-800-522-7658

Chapter 6: Maintaining Your PuriCare


8. Plug in the station and have it recertified before use.


Diffuser Removal

Bending or distorting the diffuser may
affect the airflow in the station. Use
care when handling it.


Upper Diffuser Removal

1. Swing the sash up until it is in the locked position.
2. Locate and remove the four acorn nuts that secure

the upper diffuser brackets to the station. Support
the diffuser when removing the brackets and lower
the diffuser straight down.


Lower Diffuser Removal

1. Locate and remove the six screws that secure the

lower diffuser to the station, starting with the corner
screws. Support the diffuser when removing the
last two center screws and lower the diffuser
straight down.