Diffuser removal – Labconco PuriCare Vertical Flow Stations 3830020 User Manual
Page 42
38
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
8. Plug in the station and have it recertified before use.
Diffuser Removal
Bending or distorting the diffuser may
affect the airflow in the station. Use
care when handling it.
Upper Diffuser Removal
1. Swing the sash up until it is in the locked position.
2. Locate and remove the four acorn nuts that secure
the upper diffuser brackets to the station. Support
the diffuser when removing the brackets and lower
the diffuser straight down.
Lower Diffuser Removal
1. Locate and remove the six screws that secure the
lower diffuser to the station, starting with the corner
screws. Support the diffuser when removing the
last two center screws and lower the diffuser
straight down.