Invasive blood pressure (ibp), Vitalsbridge directions for use – Laerdal VitalsBridge User Manual

Page 22

Advertising
background image

VitalsBridge Directions for Use

21

rev. B.7

PN 10229

A 2-tube cuff is configured as follows.

The tubing connection from the NIBP cuff to the patient monitor will need configuration, depending on
the style of connector used with the patient monitor’s NIBP tubing. A variety of connectors have been
included for appropriate fitting(s) to be placed at the end of the cuff tubing that connects with the
patient monitor NIBP tubing.

Caution: The NIBP cuffs have been modified to function with the VitalsBridge with the

VitalsBridge, and are not for clinical use.

Invasive Blood Pressure (IBP)

The VitalsBridge is able to simultaneously simulate up to 3 different invasive blood pressure waveforms.
These are arterial blood pressure (ABP), central Venous pressure (CVP), and pulmonary artery pressure
(PAP). The VitalsBridge simulates the transducer output signal for each invasive blood pressure and
utilizes an RJ-11 cable to connect from the VitalsBridge to the patient monitor’s IBP cable. Note: contact
Laerdal for assistance with purchasing a VitalsBridge-compatible invasive blood pressure monitoring
cable (i.e., the cable that connects from the patient monitor to the VitalsBridge RJ-11 cable).

To VitalsBridge

NIBP Tubing

To Patient

Monitor’s NIBP

Tubing

connection

Advertising