Enabling co, Simulation, Calibration – Laerdal VitalsBridge User Manual

Page 31: Vitalsbridge directions for use

Advertising
background image

VitalsBridge Directions for Use

30

rev. B.7

PN 10229

CO

2

Calibration

Calibration may be needed if the end-tidal CO

2

reading on the patient vital signs monitor does not

properly match the CO

2

waveform on the Instructor Application. From the “Calibrate” tab on the

VitalsBridge software, select “(Update)” next to the CO2 label. Select a “(New Calibration)”.

Enter the name of the custom calibration.

For each calibration point, observe the target etCO

2

value in the CO2 Calibration window. Slowly move

the indicator to adjust the flow up or down until the end-tidal CO

2

reading on the patient vital signs

monitor equals the target end-tidal CO

2

reading.