6 Operation

Version 11/2008, Rev. 2.2 of 04/12/2013 • sb

49

Translation of the original operating manual

Fig. 43: Selecting a language


Settings
The Settings menu is used to customise the operation and process
management of the control system.
• Select the value you want to set.
• Change the value using the up/down keys.
• Confirm the input by pressing the right-hand function key.

Fig. 44: Settings menu


Timer Warm-up
You can set a warm-up time between 5 and 99 minutes.

Temperature Warm-up
During the warm-up phase, the rotor temperature is set to the set
temperature.

Click at button push
If this function is activated, a brief sound signal can be heard whenever a
key is pressed.

High temperature resolution
Temperatures are displayed in the values window with a resolution of

1

/

10

°C, normally with a resolution of 1°C.


Rotor speed
The rotor speed can be set in a range between 100 rpm and 1,750 rpm.

NOTE

We guarantee that a speed of 1,350 rpm can be reached with all of the
available rotors. Higher speeds are possible in individual cases.


Continuous operation coldtrap (only if a cold trap is installed) and
Continuous operation VP
If these options are set to "yes", the vacuum pump and the cold trap will not
be switched off after the evaporation. Instead, they will remain active. This
is useful if several evaporation processes are started in succession.