7 Malfunctions and error correction

Version 11/2008, Rev. 2.2 of 04/12/2013 • sb

51

Translation of the original operating manual

Pos: 155 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 Stör ung en u nd Fehl ersuc he === == == === == == === == == === == = @ 8\m od_ 1309 939 131 742 _68. docx @ 47 129 @ 1 @ 1

7 Malfunctions and error correction

Pos: 156 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10 Allg emei ne Stö run gen @ 8\ mod _13 099 413 440 43_ 68.d ocx @ 472 24 @ 2 @ 1

7.1 General malfunctions

Malfunctions are displayed in the menu "Process & equipment info" (see
chapter 6.5.3.4 - "Process and equipment information"). A sound signal
sounds when an error message is generated.
• Eliminate the source of the problem (see the following chapters).
• Acknowledge the error messages by pressing the right-hand function

key.

Pos: 157 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 158 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 159 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10- 001 0 Fehle rm eldun gen (a uße r 2 -18 ) @ 8\m od_1 309 939 132 102 _68. docx @ 47 137 @ 3 @ 1

7.1.1 Error messages

Error

Possible reason

Correction

No vacuum sensor signal

• The control unit receives an

invalid value from the vacuum

sensor.

• Check the connecting cables

and plug-in connections of the
vacuum sensor. If necessary,

replace the vacuum sensor.

No rotor chamber sensor signal

• The temperature sensor of the

rotor chamber is disconnected

or defective

• Contact the after-sales service

(see chapter 7.2 - "Service

contact")

No rotor sensor signal

• The rotor temperature sensor is

disconnected or defective

• Contact the after-sales service

(see chapter 7.2 - "Service

contact")

No ice condenser sensor signal

• The data communication link to

the cooling trap is interrupted,
or the ice condenser
temperature sensor is

disconnected or defective

• Check the data link between the

rotational vacuum concentrator
and the cooling trap

Lid cannot be opened

• The rotor chamber is not

aerated completely

• Aerate the rotor chamber

completely

Lid contacts error

• The system has detected an

open lid during evaporation.
The rotor will be stopped
immediately. The lid must be

closed while the rotor is moving.

• Check the lid contacts. If

necessary, contact the after-
sales service (see chapter 7.2 -

"Service contact")

Rotor overtemperature

• The maximum permissible rotor

temperature of 85°C has been

exceeded

• If the system does not cool

down, disconnect it immediately
from the power supply. Contact
the after-sales service (see

chapter 7.2 - "Service contact")

Chamber overtemperature

• The maximum permissible

chamber temperature has been

exceeded

• If the system does not cool

down, disconnect it immediately
from the power supply. Contact
the after-sales service (see

chapter 7.2 - "Service contact")