8 Maintenance and service

Version 11/2008, Rev. 2.2 of 04/12/2013 • sb

59

Translation of the original operating manual

• Remove the countersunk screw with Torx

®

internal drive with the

supplied Torx

®

key (TX 15).

• Remove the thrust piece and the compression spring from the axle by

lifting it vertically upward.

• Remove the rotor shaft from the axle by lifting it vertically upward.

Abb. 55: Removing the rotor shaft


• Clean the lid with a disinfectant or cleaning agent.
• Now the rotor chamber can be cleaned with a cleaning agent or

disinfectant.

• Clean the rotor shaft and the magnetic coupling.

CAUTION

Ensure that no cleaning agent or disinfectant gets into the ball bearings of
the rotor shaft as it may wash out the ball bearing grease.
This can lead to irreparable damage.


5

Rotor shaft with

magnet coupling

6

Ball bearing

Fig. 56: Rotor shaft with ball bearings