3 autoclaving – Martin Christ RVC 2-25 CDplus User Manual

Page 62

Advertising
background image

RVC 2-25 CD&

8 Maintenance and service

62

Version 11/2008, Rev. 2.2 of 04/12/2013 • sb

Translation of the original operating manual

In order to completely remove the glass particles and metal dust from
the rotor chamber
• Grease the upper third of the rotor chamber with e.g. Vaseline.
• Then, let the rotor rotate for a few minutes at maximum speed and

atmospheric pressure. The glass and metal particles will now collect at
the greased part.

• Remove the grease with the glass and metal particles with a cloth.
• If necessary, repeat this procedure.
• Clean the rotor shaft.

Pos: 187 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 188 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /08 0 Wart ung un d Inst and haltu ng/ 080 -002 0 D esinfek tion v on R otork am mer un d Z ube hör @ 8\ mod _13 100 270 5588 9_6 8.d ocx @ 472 92 @ 2 @ 1

8.2

Disinfection of the rotor chamber and accessories

• Use commercially-available disinfectants such as, for example, Incidur

®

,

Meliseptol

®

, Sagrotan

®

, Buraton

®

or Terralin

®

(available at chemist’s

shops or drugstores).

• The rotational vacuum concentrator and the accessories consist of

various materials. A possible incompatibility must be considered.

• Before using cleaning or decontamination agents that were not

recommended by us, contact the manufacturer to ensure that such a
procedure will not damage the rotational vacuum concentrator.

• For autoclaving, consider the continuous heat resistance of the

individual materials.

• Please contact us if you have any queries (see chapter 8.4 - "Service").

DANGER

If dangerous materials (e.g. infectious and pathogenic substances) are
used, the RVC and the accessories must be disinfected.

Pos: 189 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 190 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 191 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /08 0 Wart ung un d Inst and haltu ng/ 080 -003 0 Au toklavie re n @ 8 \mo d_1 310 0272 705 88_ 68.d ocx @ 472 99 @ 2 @ 1

8.3 Autoclaving

The service life of the accessories essentially depends on the frequency of
autoclaving and use.
• Remove the existing O-rings so that the trapezoid notch can be reached

by the hot steam. This prevents the formation of humidity in the notch..

• Replace the accessories immediately when the parts show changes in

colour or structure or in the occurrence of leaks etc.

Accessories

Max. temp

[°C]

Min. time

[min]

Max. time

[min]

Max. cycles

Glass tubes

134-138

3

40

-

Teflon tubes

134-138

3

5

100

Aluminium rotors

134-138

3

20

-

Aluminium accessories

134-138

3

20

-

Pos: 192 /01 0 Univ ersal mod ule/ Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1

Advertising