7 Malfunctions and error correction

Version 11/2008, Rev. 2.2 of 04/12/2013 • sb

53

Translation of the original operating manual

Pos: 164 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10- 003 0 Syste mm eldu ngen @ 8 \mo d_1 309 9391 353 52_ 68.d ocx @ 472 01 @ 3 @ 1

7.1.3 System messages

Error

Possible reason

Correction

Invalid system type

• Contact the after-sales service

(see chapter 7.2 - "Service

contact").

IO communication error 0x21

• Failure of the CDplus IO module • Switch the system off and then

on again. If the error occurs
again, contact the after-sales
service (see chapter 7.2 -

"Service contact")

IO communication error 0x30

• Failure of the reading unit of the

electronic rotor system

• Switch the system off and then

on again. If the error occurs
again, contact the after-sales
service (see chapter 7.2 -

"Service contact")

IO communication error 0x40

• Failure of the motor control

system

• Switch the system off and then

on again. If the error occurs
again, contact the after-sales
service (see chapter 7.2 -

"Service contact")

IO communication error 0x41

Failure of the imbalance
detection module

• Switch the system off and then

on again. If the error occurs
again, contact the after-sales
service (see chapter 7.2 -

"Service contact")

Pos: 165 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 166 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 167 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 25 CD plus/0 70 Stö run gen un d F ehle rsuch e/07 0-0 010 -00 40 Note ntrie gelu ng d es Deck els @ 8\m od_ 1315 902 348 228 _68. docx @ 50 533 @ 3 @ 1

7.1.4 Emergency lid release

In the event of a power failure, it is possible to manually open the lid of the
rotational vacuum concentrator.

WARNING

• Do not actuate the emergency lid release before the rotor has stopped

and the pressure has been relieved.

• Do not use the emergency lid release to open the lid during normal

operation. Unlocked lid locks increase the risk of accidents!

<dg_ ref _sou rce_i nline >-Pro gra mmv erwal tun g-


• Switch off the mains power switch and disconnect the power cord from

the socket.

• Check if the rotor is at a standstill and the rotor chamber is completely

aerated.

• Pull the unit a bit over the edge of the table and jack it up (see the

illustration below). Ensure that the centre of gravity of the unit is still on
the table. If necessary, the unit must be secured by a second person.