8 Maintenance and service

58

Version 11/2008, Rev. 2.2 of 04/12/2013 • sb

Translation of the original operating manual

Pos: 179 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 25 CD plus/0 80 Wa rtu ng u nd I nsta ndh altun g/08 0-0 010 -00 20 R oto rkam me r @ 8\mo d_1 317 721 530 497 _68. docx @ 50 769 @ 3 @ 1

8.1.2 Rotor chamber

If the rotor chamber is contaminated or if glass is broken, the rotor
chamber, the rotor shaft and the rotor must be cleaned immediately in order
to prevent corrosion and damage to the ball bearings.

Cleaning of the rotor chamber


1

Rotor chamber

2

Cap

3

Rotor bearing

Fig. 53: Rotor chamber with rotor shaft


• Open the lid.
• Disconnect the power cord from the wall outlet or instrument receptacle

before cleaning.

DANGER

Hot surface!
Let the rotor chamber, the rotor shaft and the lid cool down to room
temperature!

• Pull off the cap.


4

Countersunk screw
with thrust piece and

compression spring

Fig. 54: Removing the countersunk screw