12 Glossary

82

Version 11/2008, Rev. 2.2 of 04/12/2013 • sb

Translation of the original operating manual

Safety pressure
Since the vacuum has a dominating influence on the product temperature,
Martin Christ Gefriertrocknungsanlagen GmbH has integrated a so-called
safety pressure function into their units in order to protect the products. If
the pressure inside the rotor chamber rises too quickly – above the safety
limit – the energy supply of the unit will be interrupted so that the
evaporation process slows down.

Fig. 61: Diagram describing the safety pressure function