RVC 2-25 CD&

7 Malfunctions and error correction

Version 11/2008, Rev. 2.2 of 04/12/2013 • sb

55

Translation of the original operating manual

Pos: 169 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10- 005 0 Kleinfl ansch ver bindu nge n @ 12\ mod_ 133 517 283 489 6_6 8.docx @ 6 288 9 @ 3 @ 1

7.1.5 Small flange connections

Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring

inside) in a centred manner between the flange connections.

• Seal the connection with the clamping ring or clamping chain by

tightening the wing nut.

• Ensure that the centring ring neither slips out of place nor gets jammed.

Fig. 49: Small flange and centring ring

Fig. 50: Small flange with centring ring

and small flange

Fig. 51: Attaching the clamping ring

Fig. 52: Tightened clamping ring