Chapter 1, Installation, Environmental conditions – Parr Instrument 6400 User Manual

Page 10: Swagelok tube fittings, Figure 1-1, Chapter 1 8, Hapter

Advertising
background image

1

Installation

P a r r I n s t r u m e n t C o m p a n y

8

c

haPter

1

Installation

Note:

Some of the following manual sections

contain information in the form of warnings,

cautions and notes that require special at-

tention. Read and follow these instructions

carefully to avoid personal injury and dam-

age to the instrument. Only personnel quali-

fied to do so, should conduct the installation

tasks described in this portion of the manual.

Environmental Conditions

The 6400 Calorimeter is completely assembled and
given a thorough test before it is shipped from the
factory. If the user follows these instructions, instal-
lation of the calorimeter should be completed with
little or no difficulty. If the factory settings are not
disturbed, only minor adjustments will be needed to
adapt the calorimeter to operating conditions in the
user’s laboratory.

This apparatus is to be used indoors. It requires at
least 4 square feet of workspace on a sturdy bench
or table in a well-ventilated area with convenient
access to an electric outlet, running water and a
drain.

Swagelok Tube Fittings

When Swagelok Tube Fittings are used, the instruc-
tions for installation are:

1. Simply insert the tubing into the Swagelok Tube

Fitting. Make sure that the tubing rests firmly
on the shoulder of the fitting and that the nut is
finger-tight.

2. Before tightening the Swagelok nut, scribe the

nut at the 6 o’clock position.

3. While holding the fitting body steady with a

back-up wrench, tighten the nut 1-1/4 turns.
Watch the scribe mark, make one complete revo-
lution and continue to the 9 o’clock position.

4. For 3/16” and 4mm or smaller tube fittings, tight-

en the Swagelok nut 3/4 turns from finger-tight.

Figure 1-1

Swagelok Tube Fittings

Swagelok tubing connections can be disconnected
and retightened many times. The same reliable leak-
proof seal can be obtained every time the connec-
tion is remade using the simple two-step procedure.

1. Insert the tubing with pre-swaged ferrules into

the fitting body until the front ferrule seats.

2. Tighten the nut by hand. Rotate the nut to the

original position with a wrench. An increase in
resistance will be encountered at the original
position. Then tighten slightly with a wrench.
Smaller tube sizes (up to 3/16” or 4mm) take less
tightening to reach the original position than
larger tube sizes.

Advertising