File management menu, Run data file manager, Menu descriptions – Parr Instrument 6400 User Manual

Page 45

Advertising
background image

Menu Descriptions

6400

4

w w w . p a r r i n s t . c o m

43

Run Data File Manager

The white upper portion of the Run Data File Manag-
er screen presents all tests in memory in a scrollable
window. Test attributes include filename (sample
ID), test type, status, and date. Touching anywhere in
the column related to a given test attribute will sort
the file list by that attribute. Successive touches will
toggle between an ascending and descending sort.

Select: This key is used to begin the file selection
process. The up/down (single arrow) and page up/
page down (double arrow) keys are used to scroll
up and down the file list. Pressing the select key
when a file is highlighted blue will highlight the file
with a cyan color. This indicates that it is selected.
Multiple files throughout the list can be selected in
this fashion.

Extend Sel: This key selects all files between the last
file selected and the file that is highlighted in blue.

Desel All: This key deselects all files previously
selected.

Rename: This key allows the user to rename the
blue highlighted filename.

Delete: This key deletes all selected files.

Convert Type: This key allows one or more selected
tests to be converted from determinations to stan-
dardizations and vice versa.

Diagnostics Menu

Provides the user with the means to test many of
the components and subsystems of the calorimeter.
These capabilities should be used in conjunction
with this instruction manual in order to obtain the
maximum benefits from these capabilities.

Pretesting Cycle: This key initiates a pretest to run
the calorimeter through the fill and cool/rinse cycles.
This function is used to pre-condition the calorim-
eter if it has been sitting idle for an extended period
of time (greater than 15 minutes).

Test Ignition Circuit: Activates the ignition circuit. A
volt meter can be placed across the firing connec-
tions to ensure that the actual firing charge is reach-
ing these contacts.

Data Logger: Displays ON/OFF status and accesses
the Data Logger Controls Menu for setting the
specific logging controls.