Table 6-2, Acid is used as a test sample, Standardizations – Parr Instrument 6400 User Manual

Page 52

Advertising
background image

6

Standardizations

P a r r I n s t r u m e n t C o m p a n y

50

Table 6-2

Calorimeter Control Limit Values in cal/g When Benzoic Acid is Used as a Test Sample

Accepted heat of combustion taken as 6318 cal/g.
Instrument precision 0.10%.
Control limits based on 99% confidence (3 sigma) values.
Values are in cal/g.

NUMBER OF

OBSERVATIONS

IN A GROUP

UCL FOR THE RANGE

(HIGH – LOW) WITHIN

THE GROUP

UCL FOR THE

RSD WITHIN

THE GROUP

MAXIMUM PERMISSIBLE

DEVIATION OF THE GROUP MEAN

FROM THE ACCEPTED VALUE OR

GRAND MEAN

1

19.0

2

23.3

0.261%

13.4

3

27.5

0.228%

10.9

4

29.7

0.209%

9.5

5

31.1

0.196%

8.5

6

32.1

0.187%

7.7

7

32.9

0.181%

7.2

8

33.5

0.175%

6.7

9

34.1

0.171%

6.3

10

34.6

0.167%

6.0

11

35.0

0.164%

5.7

12

35.3

0.161%

5.5

13

35.7

0.159%

5.3

14

36.0

0.156%

5.1

15

36.3

0.154%

4.9

16

36.5

0.153%

4.7

17

36.8

0.151%

4.6

18

37.0

0.150%

4.5

19

37.2

0.148%

4.3

20

37.4

0.147%

4.2

21

37.6

0.145%

4.1

22

37.8

0.145%

4.0

23

37.9

0.144%

4.0

24

38.1

0.143%

3.9

25

38.3

0.142%

3.8