Table 15-1, Factory default settings, Chapter 15 – Parr Instrument 6400 User Manual

Page 106: Tables, Chapter 15 104, Hapter

Advertising
background image

15

Tables

P a r r I n s t r u m e n t C o m p a n y

104

c

haPter

15

Tables

Table 15-1

Factory Default Settings

Calorimeter Operations
Operating Mode

Determination

Bomb Installed/EE

1/940.0

Heater and Pump

OFF

Operating Controls
Method of Operation

Dynamic

Reporting Units

BTU/lb

Use Spiking Correction

OFF

“OTHER” Multiplier

4.1868

Calibrate Touchscreen
LCD Backlight Timeout(s)

1200 S

LCD Backlight Intensity

70%

Print Error Messages

ON

Language

English

Spike Controls
Use Spiking

OFF

Heat of Combustion of Spike

6318.4

Use Fixed Spike

OFF

Weight of Fixed Spike

0.0

Prompt for Spike before Weight

OFF

Bomb Rinse Tank Control
Report Rinse Tank Empty

ON

Rinse Tank Capacity

150

# Rinses Left

150

Reset Rinse Tank Counter
Rinse Time

25

Rinse Flush Time

20

Clear Time

100

# of Rinse Cycles

3

Program Information and Controls
Date

XX/XX/XXXX

Time

XX:XX

Volume Level Adjust

85%

Software and Hardware Info
Settings Protect

OFF

User/Factory Settings
Feature Key
Bomb Type Select
User Function Setup
Cold Restart

User/Factory Settings
User Setup ID

64-1138

Reload Factory Default Settings
Reload User Default Settings
Save User Default Settings

Calibration Data & Controls
Calibration Run Limit

10

EE Max Std Deviation

0.0

Heat of Combustion of Standard

6318.4

Bomb Service Interval

500

Use Bomb

1

Bomb 1 Through 4
EE Value

940

Protected EE Value

OFF

Thermochemical Corrections Standardization
Fixed Fuse Correction

ON 50.0

Acid Correction

Fixed HN03, 8.0

Fixed Sulfur Correction

ON 0.0