Table 6-1, Acid is used as a test sample, Standardizations – Parr Instrument 6400 User Manual

Page 51

Advertising
background image

Standardizations

6400

6

w w w . p a r r i n s t . c o m

49

Any outliers or other tests which should not be
included in the average EE Value must be deleted
from the memory using the memory management
procedures (see Chapter 8). A list of all tests associ-
ated with any Cal ID can be printed from the Calibra-
tion Data Page using Print Standardization Runs.

The user can elect to have any number of stored
standardization runs used in determining the EE

value by entering this number on Calibration Data &
Controls Page - Calibration Run Limit.

EE Max Std Deviation on this same page establishes
the maximum allowable standard deviation for the
EE Value before an error condition is reported. The
default value is zero which turns off this limit. But
the user should enter a value appropriate for the
test being made.

Table 6-1

Calorimeter Control Limit Values in J/g When Benzoic Acid is Used as a Test Sample

Accepted heat of combustion taken as 26454 J/g.
Instrument precision 0.10%.
Control limits based on 99% confidence (3 sigma) values.
Values are in J/g.

NUMBER OF

OBSERVATIONS

IN A GROUP

UCL FOR THE RANGE

(HIGH – LOW) WITHIN

THE GROUP

UCL FOR THE

RSD WITHIN THE

GROUP

MAXIMUM PERMISSIBLE

DEVIATION OF THE GROUP MEAN

FROM THE ACCEPTED VALUE OR

GRAND MEAN

1

79.4

2

97.5

0.261%

56.1

3

115.3

0.228%

45.8

4

124.3

0.209%

39.7

5

130.1

0.196%

35.5

6

134.3

0.187%

32.4

7

137.6

0.181%

30.0

8

140.4

0.175%

28.1

9

142.7

0.171%

26.5

10

144.7

0.167%

25.1

11

146.4

0.164%

23.9

12

147.9

0.161%

22.9

13

149.4

0.159%

22.0

14

150.7

0.156%

21.2

15

151.8

0.154%

20.5

16

153.0

0.151%

19.8

17

154.0

0.151%

19.2

18

154.9

0.150%

18.7

19

155.8

0.148%

18.2

20

156.7

0.147%

17.7

21

157.4

0.146%

17.3

22

158.2

0.145%

16.9

23

158.9

0.144%

16.5

24

159.5

0.143%

16.2

25

160.2

0.142%

15.9