Hydrogen Entry Mode: This key steps through

the options for entering hydrogen content for
calculating the net heat of combustion either
manually through the touch screen or auto-
matically through the balance (USB) port.

Oxygen Entry Mode: This key steps through

the options for entering oxygen content for
calculating the net heat of combustion either
manually through the touch screen or auto-
matically through the balance (USB) port.

Nitrogen Entry Mode: This key steps through

the options for entering nitrogen content for
calculating the net heat of combustion either
manually through the touch screen or auto-
matically through the balance (USB) port.

Auto Sample ID Controls: Accesses sub-menu for
controlling the automatic assignment of sample
identification numbers.

Auto Sample ID Controls

Automatic Sample ID: When set to ON the unit

will automatically assign sample identification
numbers in accordance with parameters set by
the other three keys on this menu. When set
to OFF, the user manually enters each sample
ID when prompted to do so.

Auto Sample ID Prefix: An entry here will be

used as a prefix for all sample IDs, if the Auto-
matic Sample ID is set to ON. Press this key
to access a sub-menu for entering an alphanu-
meric prefix.

Next Auto Sample ID Number: Establishes the

initial sample number for a series of tests and
then shows the next sample ID which will be
assigned. Used when the Automatic Sample
ID is set to ON. Press this key to access a sub-
menu for entering a numeric increment.

Auto Sample ID Increment: Establishes the in-

crement between sample numbers; used when
the Automatic Sample ID is set to ON. Press
this key to access a sub-menu for entering a
numeric increment.

Sample Weight Warning Above: This key displays
and leads to a sub-menu used to set the maximum
allowable sample weight (including spike) in grams.
A warning will be given if sample weights above
this value are entered.

Spike Weight Entry Mode: This key steps through
the options for entering spike weights either manu-
ally through the touch screen, balance (USB) port or
network.

Sulfur Entry Mode: This key steps through the op-
tions for entering the sulfur correction value either
manually through the touch screen or automatically
through the balance (USB) port.

Moisture Entry Mode: This key steps through the

options for entering the moisture percentage wheth-

er manually through touch screen or automatically

through the balance (USB) port.