2. Let the operating temperature stabilize for 2 minutes.

3. Use the precision input source to provide a 15 mA input across a

channel’s input terminals.

I IN

COM

+

4. Verify that the input data is between 15110 and 15260 counts.

5. Repeat steps 3 and 4 for each remaining channel.

Verifying Voltage Inputs
(-INI4vi only)

1. Configure your module for

±

10 V input on all 4 channels using the

following configuration word values:

Address

Binary Value

Decimal Equivalent

O:e.0

0000 0000 0010 0001

33

O:e.1

0000 0000 0010 0001

33

O:e.2

0000 0000 0010 0001

33

O:e.3

0000 0000 0010 0001

33

e = slot number

2. Let the operating temperature stabilize for 2 minutes.

3. Use the precision input source to provide a 10 Vdc input across a

channel’s input terminals.

V IN

COM

+

4. Verify that the input data is between 31805 and 32131 counts.

5. Repeat steps 3 and 4 for each remaining channel.

Service

If the input data is out of range, call the Spectrum Controls Customer
Satisfaction department (206-746-9481) to make arrangements for
returning your module for factory calibration.