Installation of the equipment, 1 electrical connection, Danger – BINDER BF 53 User Manual

Page 17: Caution

Advertising
background image

BF (E1) 02/2015

page 17/51

DANGER

Explosion hazard.
Danger of death.

Do NOT operate the unit in potentially explosive areas.

KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit.

4.

Installation of the equipment

4.1 Electrical connection

Incubators with forced convection BF are supplied ready for connection. The socket must also provide a

protective conductor.

BF (230 V):
Shockproof plug, power supply voltage 230 V (1N~) +/- 10 %, 50/60 Hz

BF-UL 53, BF-UL 115, BF-UL 240, BF-UL 400 (115 V):
NEMA plug 5-15P, power supply voltage 115 V (1N~) +/- 10 %, 60 Hz

BF-UL 720 (115 V):
NEMA plug 5-20P, power supply voltage 115 V (1N~) +/- 10 %, 60 Hz

Fixed power supply connection cable with a length of 1800 mm / 70.87 in

Prior to connection and start-up, check the power supply voltage. Compare the values to the specified

data located on the unit’s type plate (unit front behind the door, bottom left-hand, chap. 1.4).

When connecting, please observe the regulations specified by the local electricity supply company and

as well as the VDE directives (for Germany). We recommend the use of a residual current circuit

breaker.

Pollution degree (acc. to IEC 61010-1): 2

Over-voltage category (acc. to IEC 61010-1): II

CAUTION

Danger of incorrect power supply voltage.
Damage to the equipment.

Check the power supply voltage before connection and start-up.

Compare the power supply voltage with the data indicated on the type plate.

See also electrical data (chap.12.4).

To completely separate the unit from the power supply, you must disconnect the power plug.

Install the unit in a way that the power plug is easily accessible and can be easily pulled in case

of danger.