3 chamber addressing – BINDER BF 53 User Manual

Page 26

Advertising
background image

BF (E1) 02/2015

page 26/51

1. Press down

button for approx. 5 seconds.

The display alternately shows “unit” and the temperature unit:

2. Press again button

.

The display alternately shows “

rASd” and the actual setting of the set-point gradient:

3. Set the desired ramp gradient with buttons

(set-point gradient in °F or °C acc. to setting

in chap. 6.4.1).

4. The set value is automatically adopted after 2 seconds.

During ramp operation the actual set-point (SPr) continually rises in accordance to the entered gradient

from the previously entered set-point to the new one (SP). The actual value follows the set-point value.
About set-point display during ramp operation see chap. 6.2.

6.4.3 Chamber addressing

If several incubators with forced convection BF are networked with a PC via the APT-COM™

communication software (option, chap. 8.2), each unit must be allocated a unique address. Addressing

takes place on the chamber controller as follows:

1. Press down

button for approx. 5 seconds.

The display alternately shows “unit” and the temperature unit:

2. Press again button

.

The display alternately shows “rASd” and the set-point gradient:

3. Press again button

.

The display alternately shows “

Adr” and the actual setting of the unit address: