5 equipment and options – BINDER BF 53 User Manual

Page 44

Advertising
background image

BF (E1) 02/2015

page 44/51

Unit size

53

115

240

400

720

Electrical data
IP system of protection acc. to EN 60529

20

20

20

20

20

Nominal voltage (±10 %) 50/60 Hz

V

230 1N~ 230 1N~ 230 1N~ 230 1N~ 230 1N~

Nominal power

kW

0.40

0.40

0.68

0.85

1.25

Energy consumption at 37 °C / 98.6 °F Wh/h

11

20

33

53

80

Unit fuse 5 x 20 mm
230V / 10A / middle-time-lag (M)

A

10

10

10

10

10

external external external external external

Power plug

shock proof plug

Installation category acc. to IEC 61010-1

II

II

II

II

II

Pollution degree acc. to IEC 61010-1

2

2

2

2

2

Electrical connection data BF-UL constructed acc. to cUL standard (for USA and Canada)

Unit size

53-UL

115-UL 240-UL 400-UL 720-UL

Electrical data
Nominal voltage (±10%) 60 Hz

V

115 1N~ 115 1N~ 115 1N~ 115 1N~ 115 1N~

Power plug

NEMA

5-15P

5-15P

5-15P

5-15P

5-20P

Nominal power

kW

0.40

0.40

0.68

0.85

1.25

Unit fuse 6,3 x 32 mm

250V / super-time-lag TT

A

12.5

12.5

12.5

12.5

16

external external external external external

Installation category acc. to IEC 61010-1

II

II

II

II

II

Pollution degree acc. to IEC 61010-1

2

2

2

2

2

Legend: 1) without outer glass door 2) up to 98 % of the set value
All technical data is specified for unloaded units with standard equipment at an ambient temperature of

+25 °C / 77 °F and a power supply voltage fluctuation of ±10. The temperature data is determined in

accordance to BINDER factory standard following DIN 12880 observing the recommended wall

clearances of 10 % of the height, width and depth of the inner chamber. Technical data refer to 100% fan

speed.
All indications are average values, typical for units produced in series. We reserve the right to

change technical specifications at any time.

If the chamber is fully loaded, the specified heating up times may vary according to the load.

12.5 Equipment and Options

To operate the incubator with forced convection BF, use only original BINDER accessories or

accessories / components from third-party suppliers authorized by BINDER. The user is

responsible for any risk arising from using unauthorized accessories.

Standard equipment
Microprocessor temperature controller with LED display and several time functions
Controller Timer functions: Delayed ON, delayed Off and temperature dependent delayed OFF
Temperature safety device class 3.1 acc. to DIN 12880 with visual temperature alarm
Adjustable ramp function
Rear exhaust duct, internal diameter 50 mm / 1.97 in with ventilation slide
Adjustable air change by rear exhaust duct (50 mm / 1.97 in ) with ventilation flap and front ventilation

slide
Four castors, 2 lockable (size 720 only)
RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer

output with RS 232/RS 422 interface converter

Advertising
This manual is related to the following products: