BINDER BF 53 User Manual

Page 20

Advertising
background image

BF (E1) 02/2015

page 20/51

5. Press

button to return to normal display (actual value display) (automatically after approx. 30

seconds).

Adjust the temperature safety device following any changes of the set-point (chap. 7).


6.2 Display / entry of temperature and ventilation set-points (with selected

temperature ramp)

If previously a temperature ramp value has been selected (chap. 6.4.2):

Press button

in normal display / actual value display during ramp operation to have displayed the

actual ramp temperature set-point changing according to the selected gradient in addition to the entered

final set-points for temperature and fan speed.
The incubator is operating, the controller is in normal display (actual value

display). The

actual temperature value (example: 21.8 °C) is displayed:

1. Press

button

The display shows alternately “

SPr” and the actual temperature ramp set-point changing according

to the selected gradient (example: 42.7 °C):

This ramp set-point is only displayed, not adjustable.

2. Press

button

The display shows alternately “

SP” and the previous temperature set-point (example: 60 °C):

3. With the

buttons enter a set-point value between 0 and 100.

The desired temperature set-point can be selected in a temperature range from 5 °C /

9 °F above room temperature up to 100 °C / 212 °F.

Wait 2

seconds until the entered temperature value

is

taken over

(display flashing once).

4. Press

button to proceed to the fan speed entry.

The display shows alternately “

n” and the previous fan speed set-point (example: 100%):

5. Set the desired fan speed with the

buttons

The fan speed can be set to a value between 0% and 100%.

Wait 2 seconds until the entered value is taken over (display flashing once).

Advertising