2 continuous operation – BINDER BF 53 User Manual

Page 22

Advertising
background image

BF (E1) 02/2015

page 22/51

6.3.1 Switching between Continuous operation and Timer operation

Press the time management button

.

The controller displays the actual time function. In time function “Continuous operation”, “t1” and “t inf” are

displayed alternately. In time function “Timer operation”, “t1” is displayed alternately with the running-down

time or “tOff”.
If in time function “Timer operation” the Timer is just running off (“t1”displayed alternately with the running-

down time) the timer must at first be set to Zero (chap. 6.3.3). Now “t1” is displayed alternately with “tOff”,

and the controller can be changed to time function “Continuous operation”.

Continuous operation

2 seconds

2 seconds

Timer operation

Timer not programmed or run-down

Press

button to return to normal display / actual value display (automatically after approx. 30 sec).

6.3.2 Continuous operation

1. Press the time management button

.The timer indicates its current time function.

2. If necessary, switch to Continuous operation by button

.

The display shows alternately “

t1” and the time function “Continuous operation” “t inf”:

3. Press

button to return to normal display (actual value display) (automatically after approx. 30

seconds).

The

actual temperature value (example: 21.8 °C) is displayed:

Now the controller operates with the entered set-points (chap. 6.1) in continuous operation. The heating is

permanently active, independent of the timer setting.
To cancel Continuous operation, proceed accordingly:

1. Press the time management button

.

2. Switch to Timer operation by pressing down button

for 2 seconds (chap. 6.3.1).

Advertising
This manual is related to the following products: