Electrical specifications – AMETEK XT Series User Manual

Page 17

Advertising
background image

Features and Specifications

Electrical Specifications

15

Electrical Specifications

Specifications are warranted over a temperature range of 0 to 30 °C with default
local sensing. Above 30 °C, derate output linearly to zero at 70 °C. Specifications are
subject to change without notice.

Table 1.2 Electrical Specifications for 7.5 V to 30 V Models

Models

7-6

15-4

20-3

30-2

Output Ratings:

Output Voltage
Output Current
Output Power

0-7 V
0-6 A
42 W

0-15 V
0-4 A
60 W

0-20 V
0-3 A
60 W

0-30 V
0-2 A
60 W

Line Regulation:

1

Voltage (0.01% of Vmax + 2 mV)
Current (0.01% of Imax + 250 uA)

1. For input voltage variation over the AC input voltage range, with constant rated load.

2.7 mV
0.85 mA

3.5 mV
0.65 mA

4 mV
0.55 mA

5 mV
0.45 mA

Load Regulation:

2

Voltage (0.01% of Vmax + 2 mV)
Current (0.01% of Imax + 250 uA)

2. For 0 to 100% load variation, with constant nominal line voltage, over 15-minute time

duration following load step.

2.7 mV
0.85 mA

3.5 mV
0.65 mA

4 mV
0.55 mA

5 mV
0.45 mA

Meter Accuracy:

Voltage (1% of Vmax + 1 count)
Current (1% of Imax + 1 count)

0.08 V
0.07 A

0.25 V
0.05 A

0.3 V
0.04 A

0.4 V
0.03 A

Stability:

3

Voltage (0.02% of Vmax)
Current (0.03% of Imax)

3. Maximum drift over 8 hours with constant line, load, and temperature, after 30-minute

warm-up at current setpoint.

1.4 mV
1.8 mA

3 mV
1.2 mA

4 mV
0.9 mA

6 mV
0.6 mA

Temperature Coefficient

4

Voltage (0.015% of Vmax/°C)
Current (0.02% of Imax/°C)

4. Change in output per °C change in ambient temperature, with constant line and load,

and after 30-minute warm-up at current setpoint.

1.05 mV
1.2 mA

2.25 mV
0.8 mA

3 mV
0.6 mA

4.5 mV
0.4 mA

Front Panel Voltage:

Control Resolution
(0.02% of V max)

1.4mV

3mV

4mV

6mV

Advertising