Instruction Manual

Appendix E Rev. 4.5
June 1999

E-6 Appendices

Rosemount Analytical Inc. A Division of Emerson Process Management

World Class 3000

YES

YES

OXYGEN ANALYZER

SYSTEM IS EQUIPPED

WITH HPS 3000.

CHECK JUMPERS JM6 ON

MICROPROCESSOR BOARD

AND JM1 ON INTERCONNECT

BOARD ARE INSTALLED.

CHECK JUMPERS 9

OR 10 ON POWER

SUPPLY BOARD FOR COR-

RECT CONFIGURATIONS.

REFER TO PROBE

TROUBLESHOOTING,

APPENDIX A, SECTION 2.

CHECK THAT JUMPERS

JM6 ON MICROPROCESSOR

BOARD AND JM1 ON

INTERCONNECT BOARD

ARE REMOVED.

REFER TO APPENDIX B,

SECTION 2, FOR

JUMPER LOCATIONS.

DISCONNECT POWER

AND INSTALL JUMPERS

CORRECTLY. REFER TO

FIGURE 2-5 OF MAIN IB

SECTION.

NO

NO

SYMPTOM - MICROPROCESSOR BOARD LED IS STEADY ON

27270021

Figure E-4. IFT Troubleshooting Flowchart, #2

END OF FLOWCHART #2

SYMPTOM - GENERAL USER INTERFACE OR

LED DISPLAY PANEL DOES NOT FUNCTION

REPLACE GUI BOARD.

STILL NOT FUNCTIONING.

CALL FACTORY FOR

ASSISTANCE.

27270020

Figure E-5. IFT Troubleshooting Flowchart, #3

Advertising