5-12 General User Interface (GUI) Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

World Class 3000

Table 5-5. SETUP Sub-Menu

SUB-MENU

SELECTION

PARAMETERS

DESCRIPTION

Calibration

High Gas

____% O

2

Value of high O

2

calibration gas (0.1000% - 25.00%

O

2

).

Low Gas

____% O

2

Value of low O

2

calibration gas (0.1000% - 25.00%

O

2

).

Auto Cal

YES/NO

MPS required for Auto Cal.

Output Tracks

YES/NO

NO, locks output during calibration.

Cal Intrvl

XD XH

Select time between calibrations in number of days
and hours (1 year max).

Next Cal

XH

Time until next calibration in number of hours
(1 year max).

Gas Time

0:30 - 20:00

Amount of time calibration gases will be turned on in
number of minutes and seconds; allow enough time
for signal values to stabilize.

Purge Time

0:30 - 20:00

Amount of time for gas lines to clear of calibration
gas.

Res Alarm

50 W – 10 kW

Resistance alarm set from 50 to 10,000 ohms.

O

2

Calculation

Slope

____ mV/D

Constant

____ mV

Set Point

____°C

Set value between 34.5 and 57.5.
Set value between -20.0 and +20.0 mV.
Set either 736 for World Class 3000 probes or 843
for 218 probes.

Ensure the correct voltage is selected when using HPS 3000 with either World
Class 3000 probes or 218 probes. Refer to Figure 2-15,
Jumper Selection Label
for proper voltage selections. If incorrect SET POINT is selected, damage to
the probe may occur.

Reset slope and constants.

Press ENTER to reset slope and constants to values
from the latest successful calibration.

O

2

Alarms

Hi Alarm

____% O

2

Lo Alarm

____% O

2

Alarm DB

____% O

2

Set value for high alarm limit (0.1000% - 25.00%).
Set value for low alarm limit (0.1000% - 25.00%).
Set value for alarm dead band (0.0000% - 25.00%).

Efficiency Calc.

Enable Calc.

YES/NO

K1 Value

_______

K2 Value

_______

K3 Value

_______

Select YES to enable, NO to disable.
Set between 0.0000 and 1.000. Refer to Table 5-6.
Set between 0.0000 and 1.000. Refer to Table 5-6.
Set between 1.000 and 20.00. Refer to Table 5-6.