Instruction Manual

Appendix E Rev. 4.5
June 1999

E-12 Appendices

Rosemount Analytical Inc. A Division of Emerson Process Management

World Class 3000

i. Replace protective cover (13) and secure

with washers (15) and screws (14).

j. Close cover door (16) and secure with

screws (17).

E-11 GUI ASSEMBLY REPLACEMENT

These replacement instructions are provided for
GUI equipped systems. Refer to Figure E-7.

Disconnect and lock out power before
working on any electrical components.
There is voltage up to 240 Vac, and
could cause personal injury.

a. Turn off power to the system.

b. Open cover door (43, Figure E-7) of the IFT

by removing screws (5).

1. Screw

2. Washer

3. GUI

Assembly

4. Cover

Door

5. Screw,

M4

6. O-Ring

ENT

ER

ESC

DATA

CAL

SET

UP

HEL

P

NOTES:

1.

2.

GUI MOUNT AREA IS PART OF COVER DOOR.
GUI FACES INSIDE OF IFT ENCLOSURE WHEN

COVER DOOR IS CLOSED.

3. NOT ALL PARTS SHOWN ARE AVAILABLE FOR

PURCHASE SEPARATELY. FOR A LIST OF

AVAILABLE PARTS, SEE TABLE E-3.

(SEE

NOTE 1)

1

2

3

4

5

6

21240001

Figure E-7. Replacing the GUI Assembly

c. Disconnect GUI assembly cables from mi-

croprocessor board (11, Figure E-6).

d. Remove GUI assembly (3, Figure E-7) by

removing screws (1) and washers (2).

e. Attach new GUI assembly to inside of cover

door (4) with washers (2) and screws (1).

f. Reconnect GUI assembly cables to micro-

processor board (11, Figure E-6).

g. Close cover door (4, Figure E-7) and secure

with screws (5).

E-12 HEATER / FAN / THERMOSWITCH

REPLACEMENT

These replacement instructions are provided for
heater/fan/thermoswitch equipped systems.

a. Heater

Replacement

Disconnect and lock out power before
working on any electrical components.
There is voltage up to 240 Vac, and
could cause personal injury.

1. Turn off power to the system.

2. Open cover door (1, Figure E-8) of the

IFT by removing screws (5, Figure
E-7).

3. Remove protective cover (4) by re-

moving screws (2) and washers (3).

4. Disconnect cable (1, Figure E-6) from

microprocessor board (11, Figure E-6).

5. Partially remove mounting plate as-

sembly (15) from enclosure (23) by
removing necessary screws.

6. Remove screws (11) and washers (12)

to free heater mounting plate (20) from
mounting plate (15).

7. Cut wires to old heater (21) close to the

heater; then remove old heater from
heater mounting plate (20).

Advertising